Agilent Technologies XX3625 Service Manual page 47

Table of Contents

Advertisement

Scale
2 V/div*
1 V/div
500 mV
200 mV
100 mV
50 mV
20 mV
10 mV
* only in 1 M Ω input
Figure 3-6
Source Selection for Vavg Measurement
Offset
Supply
Tolerance
± 163.8 V
2.5 V
5 V
± 82 mV
2.5 V
5 V
± 41 mV
1.75 V
3.5 V
± 16.4 mV
700 mV
1.4 V
± 8.2 mV
350 mV
700 mV
± 4.1 mV
175 mV
350 mV
± 1.64 mV
70 mV
140 mV
± 0.819 mV
35 mV
70 mV
Select Channel 1 as the
source for the Vavg
measurement
Chapter 3: Testing Performance
To test voltage measurement accuracy
Limits
4.836 V to 5.164 V
4.918 to 5.082 V
3.459 to 3.541 V
1.384 to 1.416 V
691.8 mV to 708.2 mV
345.9 mV to 354.1 mV
138.36 mV to 141.64 mV
69.18 mV to 70.82 mV