HP 8566B Adjustment Manual page 99

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

17. Image, Multiple, and Out of Rand Responses Test
2-74 Performance Tests
1. Connect equipment as shown in Figure 2-31 with the synthesized
sweeper RF OUTPUT connected to the analyzer input.
2. Press INSTR PRESET on the synthesized sweeper. Key in the
following sweeper settings:
CW FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3000.000 MHz
POWER LEVEL
3. Press (2) on the analyzer. Key in the following analyzer
settings:
[CENTER FREQUENCY) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
[
FREQUENCY
4. Press DISPLAY LINE (ENTER) and key in -70 dBm.
5. Press MARKER
disappear from the display. Press MARKER (j).
6. Adjust the power level of the synthesized sweeper to place the
peak of the signal trace at the top CRT graticule line.
If the maximum output power level of the synthesized sweeper is not
enough to place
the
signal peak at the top CRT graticule line, then
adjust the spectrum analyzer REFERENCE LEVEL as.required to
place the signal peak at the top CRT graticule line.
7. Press MARKER Ia]. Using the DATA knob, determine the
amplitude of the spurious response and enter the result in
8. Set the synthesized sweeper to the frequencies in 'Pable 2-20
corresponding to an analyzer center frequency of 3 GHz. The
maximum allowable amplitude of the spurious response at the
analyzer center frequency for each setting is shown in the table.
)
and wait for PEAKING! message to
PEAK
3 GHz

Advertisement