HP 8566B Adjustment Manual page 325

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

sweeper front panel LEVELING EXT INPUT using a BNC to SMB
snap-on test cable.
Frequency Response Adjustments Setup (2.0 GHz to 22.0
54. Press (2-22j on the synthesized sweeper. Set the controls of
the synthesized sweeper as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . -4.0 dBm
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .INT
55. On the synthesized sweeper, press
ENTRY knob for a power meter indication of -10.00 dBm f0.03
ENTRY knob (REF in dBV with ATN: 0 dB) for a power meter
indication of -10.00 dBm *0.03 dB at 100 MHz.
Do not vary the synthesized sweeper POWER LEVEL setting
(internal leveling) or XTAL REF and XTAL ATN settings (external
diode detector leveling) for the remaining steps in this adjustment
procedure. The frequency response adjustments are referenced to
-10.00 dBm power level at 100 MHz.
56. Disconnect the power sensor from the power splitter, and
connect this power splitter resistive output to the spectrum
analyzer RF INPUT using an APC 3.5 (m) to Type N (m) adapter.
57. On the synthesized sweeper, key in Icw) 5.7 GHz.
YTX Drive adjustments A6A12R63 5.8 GHz and A6A12R66 2 GHz are
interactive and affect YTX/YTO tracking in all preselected frequency
bands (Bands B, C, D, and E, 2 GHz - 22 GHz).
58. Adjust 25-turn (YTX Drive) potentiometer A6A12R63 5.8 GHz
and 17-turn (Band B YTX Diode Bias) potentiometer A6AlOR18
VB as necessary to maximize the amplitude of
2 1. Frequency Response Adjustments
Figure 3-94.
]
and adjust the
LEVEL
the 5.7
Adjustments 3-185
the

Advertisement