HP 8566B Adjustment Manual page 105

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

18. Gain Compression Test
30. Connect cable to power sensor and adjust output level of
synthesizer for a power meter indication of -5.00 ho.02 dB.
Reconnect cable to spectrum analyzer input.
[
31. Change
REFERENCE LEVEL
Record the MKR A amplitude:
32. Subtract the value recorded in step 9 from the value obtained
in step 31 to find the gain compression:
result should be >- 1.0 dB (less than 1 dB compression).
33. Disconnect cable from the spectrum analyzer RF INPUT. Connect
the spectrum analyzer CAL OUTPUT to RF INPUT.
34. Press c-1 @, MARKER
for a MARKER amplitude of -10.00 dBm f0.02 dB.
)
to -5 dBm.
(P
).
Adjust AMPTD CAL
EAK SEARCH