HP 8566B Adjustment Manual page 154

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

Reference Adjustment Range of Values
Designator Procedure
3-14 Adjustments
1.78 K to 13.3 K
8.25 K to 82.5 K
1.0 K to 6.81 K
1 K-OPEN
8
1.0 to 8.2
8
180 to 270
8
1.0 to 8.2
8
1.0 to 8.2
8
180 to 270
8
180 to 270
8
180 to 270
0 to 9.09
3.16 K to 8.25 K
6.19 K to 12.1 K
383 to 825
6.19 K to 12.1 K
1 K to OPEN
1 K to OPEN
0 to 100
3.1 6 K to 8.25 K Adjusts LC filter bandwidth.
6.19 K to 12.1 K
909 to 2.73 K
100 K to 1M
O-16
10
11
1.62 K to 2.61 K
10
1.33 K to 3.48 K
10
472 to 1.62 K
10
215 to OPEN
42.2 to 75.0
56 to 82
7
56 to 82
7
56 to 82
7
56 to 82
7
56 to 82
7
1.5 to 12.0
10.0 K to 17.8 K
10.0 K to 17.8 K
10.0 K to 17.8 K
10.0 K to 17.8 K
Function of Component
Log fidelity
Sets adjustment range of SYM.
Sets adjustment range of LC CTR.
Sets adjustment range of SYM.
Sets adjustment range of SYM.
Sets adjustment range of LC CTR.
Sets adjustment range of LC CTR.
Matches amplitude of LC to XTAL bandwidths.
Adjusts LC filter bandwidth.
Adjusts crystal filter bandwidth.
Matches amplitude of LC to XTAL bandwidths.
Adjusts crystal filter bandwidth.
Sets level of + 10 V TC supply.
Sets level of + 10 V TC supply.
Adjusts bandwidth shape in 10 kHz bandwidth.
Adjusts crystal filter bandwidth.
Adjusts positive feedback.
Sets adjustment range of LC amplitudes.
Sets adjustment range of FREQ ZERO COARSE.
Sets 18.4 MHz Local Oscillator power.
Adjusts A8dB step.
Adjust A4dB step.
Adjusts A2dB step.
Adjusts level of 3 MHz output.
Centers first pole.
Sets adjustment range of second pole P K.
Sets adjustment range of third pole P K.
Sets adjustment range of fourth pole P K.
Sets adjustment range of fifth pole P K.
Centers first pole.
Adjusts crystal filter bandwidth.
Adjusts crystal filter bandwidth.
Adjusts crystal filter bandwidth.
Adjusts crystal filter bandwidth.

Advertisement

Table of Contents
loading

Table of Contents