Center Frequency Readout Accuracy Test - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

1. Center Frequency Readout Accuracy Test

A synthesized signal source that is phase-locked to a known frequency
standard is used to input a signal to the analyzer. The frequency
readout of the analyzer is compared to the actual input frequency
for several different frequency settings over the analyzer's range.
The signal source is phase-locked to a standard known to be as
accurate as the analyzer's internal frequency reference to minimize
the "frequency reference error X center frequency" term of the
specifications.
Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8340A
Frequency Standard . . . . . a 10 MHz standard with accuracy within
Adapter, Type N (m) to SMA (f) , . . . . . . . . . . . . . . . . . . . . . 1250-1250
Adapter, SMA (f) to SMA (f) . . . . . . . . . . . . . . . . . . . . . . .1250-1158
Cable Assembly, SMA Male Connectors . . . . . . . . . . . . . . . . . .5061-1086
1.
Connect CAL OUTPUT to RF INPUT.
2.
Press c$???KJ IRECALL) @.
3.
Adjust FREQ ZERO for a maximum amplitude trace.
4.
Press (2-j.
5.
Set the synthesized sweeper for a 2.000000 GHz signal at a level of
approximately 0 dBm.
6.
Connect equipment as shown in Figure 2-l.
7.
Set analyzer
synthesized sweeper frequency according to Table 2-2. At each
setting, press
SEARCH
Adjust
LEVEL
convenient level.
8.
Record the CENTER FREQUENCY readout in the table for each
setting. The limits for this frequency are given in the table. Refer
to Figure 2-2.
)
and
(
FREQUENCY
FREQUENCY
],
)
as necessary to place signal peak at a
and
Performance Tests 2-5

Advertisement

Table of Contents
loading

Table of Contents