Frequency Response Test Setup (60 MHz to 2.5 GHz, 2 to 22 GEIz)
36. Connect equipment as shown in Figure 2-19, with one resistive
output of the power splitter connected to the power meter/power
sensor, and the second resistive output connected to the spectrum
analyzer RF INPUT using an APC 3.5 (m) to Type N (m) adapter.
Connect the power meter rear panel RECORDER OUTPUT to the
synthesized sweeper front panel LEVELING EXT INPUT.
37. On the power meter, verify that the
Consult the power sensor Cal Factor versus Frequency graph or
table and set the power meter CAL FACTOR % switch to the 100
MHz calibration setting.
38. Press
]
on the synthesized sweeper. Set the controls
PRESET
of the synthesized sweeper as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . -9.0 dBm
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . INT
39. On the synthesized sweeper, press
ENTRY knob for a power meter indication of -15.00 dBm fO.10
40. On the power meter, press
On
41.
the synthesized sweeper, press
ENTRY knob for a power meter indication of -10.00 ho.03 dB at
100 MHz.
42. On the synthesized sweeper, press (meter_) leveling and adjust the
ENTRY knob (REF in dBV with ATN: 0 dB) for a power meter
indication of -10.00 dBm f0.03 dB at 100 MHz.
Do not vary the synthesized sweeper POWER LEVEL setting (internal
leveling) or METER REF and METER ATN settings (external power
meter leveling) for the remaining steps in this test procedure.
10. Frequency Response Test
Figure 2-19.
[
)
RANGE HOLD
(
and adjust the
POWER
(
)
(turning it on).
RANGE HOLD
[
)
and adjust the
POWER LEVEL
Performance Tests 2-39
switch is off.