Steps 23 through 28.
Min
Measured Max
Steps
23. Recorded deviations from step 12.
Largest Positive 0 to -70 dBm
Largest Negative 0 to -70 dBm
Largest Positive -80 to -120 dBm
Largest Negative -80 to -120 dBm
24. Recorded deviation from steps 18 and 22.
Largest Positive step 18
Largest Negative step 18
Largest Positive step 22
Largest Negative step 22
25.
0.6 dB
Sum of Positive Deviations of steps 23 and 24
26.
-0.6 dB
Sum of Negative Deviations of steps 23 and 24
27.
1.0 dB
Sum of Positive Deviations of steps 23 and 24
28.
-1.0 dB
Sum of Positive Deviations of steps 23 and 24
2.99 Performance Tests