Using The Agilent E6175A 8-Channel Load Card - Agilent Technologies E6198B User Manual

Functional test system switch/load unit
Table of Contents

Advertisement

Using the Agilent E6175A 8-Channel Load Card

E6198B Switch/Load Unit User Manual
The Agilent E6175A 8-channel high-current load card, with
current sensing, is intended to be used with loads mounted
inside the Switch/Load Unit. This card provides great flexibility,
low series resistance, and high current-carrying capability.
Figure 5-3
shows a block diagram of the Agilent E6175A
8-Channel Load Card. The card layout is shown in
• The card is two slots wide (4 cm./1.6 inches) to allow
mounting of larger loads. Due to its high current capability, it
is recommended that you mount it on the left side (slots 1-4)
of the Agilent E6198B Switch/Load Unit enclosure, closest to
the incoming power buses.
Recommendations"
• The card type is 01
• The factory default is to load a 0.05 , 0.1% Isense resistor in
each channel. If you prefer to use a current transducer, you
must remove the Isense resistor and add the current
transducer described in
• A nine-inch by four-inch area of sheet metal is left open on
the front of the card for mounting loads. You provide the load
mounting hardware, drilling holes in the sheet metal as
needed.
"Connecting
• Each channel is capable of up to 7.5 amperes continuous
carry current or up to 15 amperes with a two percent duty
cycle.
• There is a slow blow fuse on each channel to protect the card
traces against extended high current operation.
Using Load Cards and Loads
"Card Location
.
"Load Card Type and Configuration
h
"Selecting a Current-Sense
Loads".
5
Figure
5-4.
ID".
Method".
5-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ts-5000

Table of Contents