Error Messages

œ Errors are retrieved in first-in-first-out (
returned is the first error that was stored. When you have read all
errors from the queue, the
generator beeps once each time an error is generated.
œ If more than 20 errors have occurred, the last error stored in the
queue (the most recent error) is replaced with -350, "Too many errors".
No additional errors are stored until you remove errors from the
queue. If no errors have occurred when you read the error queue,
the function generator responds with +0, "No error".
œ The error queue is cleared when power has been off or after a *CLS
(clear status) command has been executed. The *RST (reset)
command does not clear the error queue.
œ Front-Panel Operation:
3: ERROR
(SYS MENU)
If the
annunciator is on, press
ERROR
the errors stored in the queue. The errors are listed horizontally on
the "
" level. All errors are cleared when you go to the
PARAMETER
"
" level and then turn off the menu.
PARAMETER
First error in queue
œ Remote Interface Operation:
SYSTem:ERRor?
Errors have the following format (the error string may contain
up to 80 characters):
-113,"Undefined header"
228
FIFO
annunciator turns off. The function
ERROR
Shift
ERR 1:
-113
Reads one error from the error queue
) order. The first error
(Recall Menu) to read
<
Error code