Chapter 4 Command Reference
Introduction
Comments
When :TRIG:COUN:AUTO is ON and :CAL3:AVER[:STAT] is ON, the
Counter clears the statistics results and the statistics current count on
:INIT[:IMM].
If the instrument is already in the process of making a measurement or if
INITiate:CONTinuous is set to ON, an :IMMediate command has no affect,
and an error -213
(Init ignored) is generated.
When :CALC2:LIM:CLE:AUTO is ON, the Counter performs the following
whenever the :INIT[:IMM] command is sent:
– invalidates the limit data,
– clears :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt information,
– turns off the front-panel display' s Limit annunciator, and
– sets the Limit-Detect output to the in-limit voltage level.
This command is an overlapped command (see IEEE 488.2, Section 12).
Beginning a measurement or block of measurements with an
:INITiate[:IMMediate] sets the Pending Operation Flag to true. Completing
the measurement or block of measurements (normally or by aborting) sets
Pending Operation Flag to false.
For an automatically armed Totalize measurement, use :ABORt to terminate
the measurement initiated with :INIT[:IMM].
When [:SENS]:EVEN:LEV[:ABS]:AUTO is ON, the Counter performs an
auto-trigger on the measurement channel(s) whenever the :INIT[:IMM]
command is executed.
Related
Stop/Single
Front-Panel
Keys
Programming Guide
4-47