Chapter 4 Command Reference
Introduction
:CALCulate2:LIMit:PCOunt[:TOTal]?
Queries the total Pass COunt (that is, the number of measurements that passed the
limit test).
Query Response
Comments
Query only.
:CALCulate2:LIMit:STATe <Boolean>
Sets or queries the limit test enable .
Query Response
Comments
Related
Limit Modes
Front-Panel
Keys
Programming Guide
Numerical data transferred as ASCII bytes in <NR1> format.
If CALC2:LIM:STATe is OFF, 0 is returned and error -221 is generated.
If no valid result exists, 0 is returned and error -230 is generated.
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
When :CALC2:LIM:STAT OFF is sent, it causes the Counter to:
– Invalidate the limit data and clear the information in
:CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt.
– Turn off the front-panel display' s Limit annunciator.
– Set the Limit-Detect output to the in-limit voltage level.