3.5 Wiring method

Caution on wiring
CAUTION
Each external terminal is provided on the rear panel of the analyzer. (See Fig. 3-3)
Wire each terminal, referring to Fig. 3-3 and (1) to (7).
(1) Power inlet
When using supplied power cable, connect the female side to the power inlet at the rear panel of the
analyzer, and insert the male side into a receptacle matching the rating.
Teledyne Analytical Instruments
• Wiring work must be performed with the main
power set to OFF to prevent electric shocks.
• Enforce construction of class-D grounding wire
by all means.
If the specified grounding construction is
neglected, a shock hazard or fault may be caused.
• Wires should be the proper one meeting the
ratings of this instrument. If using a wire which
cannot endure the ratings, a fire may occur.
• Be sure to use a power supply of correct rating.
Connection of power supply of incorrect rating may
cause fire.
Fig. 3-3 Rear panel
15
Model 7500 Instruction Manual