Password Protection - Teledyne 4040 Instruction, Operating And Maintenance Manual

Methane/ nonmethane analyzer
Table of Contents

Advertisement

Operation
To setup the
GAIN
1) From the Analysis Screen, press an arrow key to enter the Setup
Mode. The VFD will display the first 2 lines of functions available.
________________________________________________
-> GAIN ADJUST
PASSWORD
________________________________________________
<▲▼>
If the arrow is not adjacent to the GAIN ADJUST menu
item, use the ▲▼ keys to move the arrow to the proper
position.
<Enter>
Press <Enter> to activate the function and move you to
the next screen.
________________________________________________
-> GAIN 1
________________________________________________
Note: For instruction on how to, please refer to the section 4.6.3
Navigation and Data Entry.
In the example, the number to the right of the word 'GAIN' is 1,
this reflects a gain of x10.

4.8 Password Protection

Before a unique password is assigned, the system assigns TAI by
default. If the password is the default, it will be displayed automatically.
The operator just presses <Enter> from the any of the MAIN MENU
items to be allowed access to that item's sub menu. If the password has
previously been changed from the default then the initial display will be
'A' 'A' 'A', and the correct letters must be input.
If user has LOGGED OUT, then only the following SUB MENU
procedures may be executed:
14. PASSWORD
15. LOGOUT (which only repeats the logout)
16. MODEL
Teledyne Analytical Instruments
54
:
Series 4040

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents