Advanced User Functions - Teledyne 4030 Series Instruction Manual

Total hydrocarbon (thc) analyzer
Table of Contents

Advertisement

'Gain: ' plus the Offset value plus a value that is proportionate to the
number indicated to the right of 'GAIN:'.
If these 2 values are modified and entered and accepted, then the
analog output will be affected in like manner in ANALYSIS MODE.
That is, it will be scaled by the gain value, then summed with the offset
value.
The analog output for each channel may be set to source THC, SMP,
AVG, WAVE, or R-ID. In the case of THC/SMP/AVG, the output will
reflect the percent of range for the respective selection. In the case of
WAVE, the output will track the FID output. In the case of R-ID, the
output will reflect the selected range: LOW: 1/5 of full scale, MID: 1/2
of full scale, HIGH: 4/5 of full scale.
4.4.15 Standby
This function allows you to place the instrument in STANDBY.
CAUTION:
STANDY SHUTS DOWN POWER TO THE DISPLAYS
ONLY. INTERNAL CIRCUITS ARE STILL ENERGIZED
AND ELECTRICAL SHOCK HAZARD STILL EXISTS.
To place the instrument in STANDBY status:
<▲▼>
From the MAIN MENU, scroll to the STANDBY
function using the UP/DOWN keys.
<Enter>
Pressing <Enter> places the instrument in STANDBY.
To exit STANDBY, scroll again to the STANDBY
function and press <Enter> again.

4.5 Advanced User Functions

The Model 4030 provides additional functions for tailoring the
instrument to your specific application. These functions include:
20. Timing
4.5.1 Timing
Timing controls the parameters of the valve and data acquisition. It
is specific to the setup of the instrument and must be determined at the
factory or TAI trained personnel.
Teledyne Analytical Instruments
Operation
52

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents