4.6.5.2 ESCAPE

Pressing the ESCAPE key takes you back to the previous screen.
If you do not wish to continue a function, you can abort the session
by escaping. Escaping a function takes the analyzer back to the previous
screen, or to the ANALYZE Function, depending on the nesting level of
screens in the function you are moving out of.

4.6.6 Setting up an AUTO-CAL

If the analyzer is equipped with the optional auto-calibration feature
(see Chapter 3,
Installation),
sequence of steps that automatically zero and span the instrument (auto
calibration or autocal).
Note: Before setting up an AUTOCAL, be sure you understand
the Zero and Span functions as described in Section
4.6.14, and follow the precautions given there.
Note: If you require highly accurate AUTOCAL timing, use
external AUTOCAL control where possible. The internal
clock in the Model 4020 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.
Note: By default, AUTOCAL is off.
To setup an
AUTOCAL
certain number of days:
<any key>
Press any key to enter the Setup Mode. The VFD will
display the first 2 lines of functions available.
<▲▼>
If the arrow is not adjacent to the AUTOCAL menu item,
use the ▲▼ keys to move the arrow to the proper
position.
<Enter>
Press <Enter> to activate the function and move you to
the next screen.
Teledyne Analytical Instruments
the analyzer can cycle itself through a
cycle for a zero or span event to occur in a
Model 4020
40