CHARGE BATT

6.3.3 Normal Operator Induced Messages

M
ESSAGE
DOING MANUAL ZERO
CALIBRATION
DOING MANUAL SPAN
CALIBRATION
?ARE YOU SURE?
| KEEP | LOSE
[KEPT] SETPOINT
CHANGE
!LOST! SETPOINT
CHANGE
SETUP: CALGAS
SETUP: CALTIME
66
Teledyne Analytical Instruments., - Rev. 3
Either the battery should be discharged or the instrument
turned off until charging is possible.
Continued operation in BATTERY LOW mode will
cause the CARGE BATT mode listed below.
The battery in the portable model is discharged.
The instrument does not function as an analyzer while in
this mode.
The instrument should be turned off and the battery
charged before further operation.
Failure to turn off the instrument while in CHARGE
BATT mode can shorten the life of the battery.
A manual zero calibration has been requested. The
instrument is currently performing the calibration.
A manual span calibration has been requested. The
instrument is currently performing the calibration.
Prompts for saving information into one of the SETUP
submenus.
Pressing the UP key will store the changed number.
Pressing the DOWN key or allowing the 60-second
timer to expire will discard the changes and restore the
previous value.
Confirmation that a change made in one of the SETUP
submenus was stored.
Indicates that a change entered into one of the SETUP
submenus was not saved. The previous value was
restored.
Screen from the SETUP menu. It indicates the entered
value for the concentration of the span gas.
Screen from the SETUP menu. It indicates the period
between automatic calibrations.
Change the period using the front panel UP and DOWN
keys.
A value of 0 turns the AUTO calibration feature OFF.