Analog Output - Teledyne IR7000 Operating Instructions Manual

Ndir gas analyzer
Table of Contents

Advertisement

Installation 3
Relay outputs (optional)
Digital I/O (optional)
RS-232 Communications port
AC power connections

3.6.1 Analog Output

The standard model IR7000 is equipped with a single set of analog
output terminals accessible from the rear panel. The output is set at the
factory to 0–1 VDC but can be changed to 0–5V or 0–10V full scale by
moving the slides on the switch labeled "S1" on the I/O board. The switch
is located at the upper edge of the I/O board as shown in Figure 3-6. Use
the table in the figure to set the slides for the desired output voltage. A 4–
20 mA isolated or non-isolated output is also available as an option.
A second set of 0–1, 0-5, or 0-10 VDC output terminals may be
driven if the optional O
is equipped with a dual bench, there will be 2 sets of terminals, one for
each bench.
22
Teledyne Analytical Instruments., - Rev. 3
Figure 3-5: IR7000 Rear Panel
monitoring channel is installed. If the instrument
2
IR7000

Advertisement

Table of Contents
loading

Table of Contents