Rs-232 Cable - Teledyne IR7000 Operating Instructions Manual

Ndir gas analyzer
Table of Contents

Advertisement

Installation 3
P
IN
F
UNCTION
#
1
Range Bit 0 +
3
Range Bit 1 +
5
Doing Cal +
7
Alarm/Limit 1 +
9
Alarm/Limit 2 +
11
Audible Alarm +
13
Spare1+
15
Spare2 +
17
Not Used
19
Cal Req. Full/Zero +
21
Cal Req. Span +
23
Fault +
Table 3-2: Digital I/O Connector Pin Out
C
R
Installation involves:
Fabricate the I/O cable with mating connector.
Mate cable and connector to the rear panel.
Connect opposite end(s) of cable to the input and output devices.

3.6.6 RS-232 Cable

An optional serial port is available for communication to and from a
remote computer. The port is a standard RS-232 serial communications
port and allows the user to input control functions and output data in
response to a variety of requests. A standard DB-9 male connector is
mounted on the rear panel for connection to a remote PC computer via a
28
Teledyne Analytical Instruments., - Rev. 3
P
IN
F
#
2
Range Bit 0 -
4
Range Bit 1 -
6
Doing Cal -
8
Alarm/Limit 1 -
10
Alarm/Limit 2 -
12
Audible Alarm -
14
Spare1 -
16
Spare2 -
18
Not Used
20
Cal Req. Full/Zero -
22
Cal Req. Span -
24
Fault -
HART
BIT 0
ANGE
1
Off
2
On
3
Off
4
On
Table 3-3: Range ID Logic Table
UNCTION
BIT 1
Off
Off
On
On
IR7000
IN/OUT
Output
Output
Output
Output
Output
Output
Input
Input
Input
Input
Output

Advertisement

Table of Contents
loading