Main Board - Teledyne IR7000 Operating Instructions Manual

Ndir gas analyzer
Table of Contents

Advertisement

Operational Theory 2

2.3.3 Main Board

In effect, the main board imports an analog signal from the
preamplifier and outputs a digital signal. A lot of signal conditioning and
processing is performed along the way. Major functions of this board
include:
Amplification
The signal from the detector is amplified
Filter
The analog signal is filtered and conditioned
A to D Converter
The analog signal is digitized using a voltage to frequency
converter
Microprocessor
Encodes both the amplitude and phase of the digital signal
Counts, integrates, and stores the signal
Handles input and output to and from the main board
Linearizer
Scales and linearizes the signal using data and algorithms
permanently stored in the microprocessor
Filter
De-spikes, pre-filters and filters the signal again with a
filter rate chosen by the operator
The main board receives the raw signal from the detector and
amplifies it. In the analog circuit portion of the main board, the signal is
filtered to remove any electrical interference before passing it along to the
digital section as a relatively clean sine wave of several volts.
The sine wave is digitized using an onboard voltage to frequency
converter. In this process, both the amplitude and phase of the digital
signal are encoded and integrated. The microprocessor counts the digital
pulses and linearizes it using a 7th order polynomial whose coefficients
were determined at the factory based on the particular detector.
12
Teledyne Analytical Instruments., - Rev. 3
IR7000

Advertisement

Table of Contents
loading

Table of Contents