3.8 Watchdog

The AX300-I is equipped with a watchdog circuit that continuously analyzes the
electronics for proper operation. If the watchdog detects a failure, one of the codes in
Table 3.3 will appear on the LCD.
The error codes can appear on the LCD when batteries are first installed,
during normal operation or if the unit is subjected to shock. In some cases, an
additional digit is used in the error code to supply additional information. For
example, the error code 6 and 7 are followed by another digit listed as (N) in Ta-
ble 3.2. The error code 65 would indicate that a key is stuck and this key is the
Silence key.
In addition to supplying visual error codes, the audio device will beep a
number of times to indicate the general error in case the display is not functional.
Note:
To reset the watchdog error code, remove one battery for 5 seconds and re-
place. If the error persists, contact your local representative or Teledyne for as-
sistance.

Table 3.2 Error Codes

Error Code
Audio Beeps
Indeterminate
2
30
3
50
5
6 (5)
6
(There is no
indication of
which key is
stuck)
Teledyne Analytical Instruments
Error
The watchdog timer has timed out indicating
a serious software error
Analog output is different from the expected
value. May indicate a shorted or over-loaded
analog output or a failure of the analog to
digital converter or digital to analog conver-
ter circuit.
The ADC circuit failed during POST.
A stuck key has been detected. The second
digit example (5) on the display shows
which key is stuck:
1 - Batt
2 - Cal
5 - Silence
8 - Key Lock
Service Instructions
23