The Patient-Leakage-Current test is applicable only to DUTs with applied
parts.
Action
1. Position the
2. Position the
3. Set the
NEUTRAL SWITCH
4. Set the
EARTH SWITCH
5. Set the
POLARITY SWITCH
6. Connect the banana plug end of the RED test lead into the RED signal jack, and the
probe end of the RED test lead to the DUT enclosure; connect the DUT applied parts
to the applied-parts jacks on the ESA601; connect the DUT power cord to the test
receptacle on the ESA601.
7. After the meter settles, read the display, and/or press the
reading.
8. Set the
NEUTRAL SWITCH
9. After the meter settles, read the display, and/or press the
reading.
10. Set the
NEUTRAL SWITCH
11. Set the
EARTH SWITCH
12. After the meter settles, read the display, and/or press the
reading.
13. Set the
EARTH SWITCH
14. Set the
POLARITY SWITCH
to illuminate.
15. Repeat steps 8 through 13.
16. Cycle the
APPLIED
number of leads that are connected), repeating steps 8 through 15 for each applied-
parts selection.
Note
-
FUNCTION
SELECTION KNOB
-
-
APPLIED
PARTS
SELECTION KNOB
to
CLOSED
to
.
CLOSED
to
NORMAL
to
.
OPEN
to
CLOSED
to
.
OPEN
to
.
CLOSED
to
REVERSE
-
-
PARTS
SELECTION KNOB
Performing Electrical-Safety Tests
to
(
PATIENT
LEAD TO GROUND
to
.
ALL
.
, and wait for the analyzer to switch to normal.
PRINT SWITCH
PRINT SWITCH
.
PRINT SWITCH
, and wait for the
REVERSE POLARITY
from
(
1) through
RA
AP
Local Control
).
to print the
to print the
to print the
indicator
6 (
10) (or the
V
AP
5
5-19