VDE: Differential-Leakage-Current Test

The VDE: Differential-Leakage-Current test measures the magnitudes of the differential
current flowing in the Equipment Outlet L1 and L2 circuits. Figure 5-11 shows the
electrical connections between the ESA601 and the DUT. The reading is taken after
selecting this function. DUT power is ON.
S5
L1
MAINS
L2
Action
1. Position the
2. Connect the DUT power cord to the test receptacle on the ESA601.
3. After the meter settles, read the display, and/or press the
5-28
L2
OPEN
S1
S2
PE
Figure 5-11. VDE: Differential-Leakage-Current Test Configuration
-
FUNCTION
SELECTION KNOB
reading.
DUT_L1
DUT_L2
EARTH
OPEN
DUT_PE
S3
to
VDE
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
:
.
DIFF CURRENT
to print the
PRINT SWITCH
FE
ayr15f.eps