Performance Specifications - Fluke ESA601 Operator's Manual

Electrical safety analyzer
Table of Contents

Advertisement

ESA602
Operators Manual

Performance Specifications

-
LEAKAGE
CURRENT MEASUREMENT
VOLTAGE MEASUREMENT
-
EARTH
RESISTANCE MEASUREMENT
INSULATION MEASURMENT
VDE DIFFERENTIAL CURRENT
MAINS ON APPLIED PARTS
A-2
:
Ranging: Auto
Configuration: RMS current flowing through the IEC 60601-1 test load or
ANSI / AAMI ES1 test load (selectable).
Measurement Range: 0 - 8000 A True RMS
Display Units: A
Display Resolution: 1 A
±1% of reading ( ±2 µA) @ DC and 25 Hz to 1000 Hz †
Accuracy:
±2.5% of reading (±2 µA) 1kHz tp 200 kHz *†
Frequency response DC to 1 MHz (-3dB)†
†Accuracy (Mains-On-Applied-Part Leakage Current): ± 2 % of
reading, ± 6 µA
Crest Factor: <3
Input Impedance: Per Figure 15 of IEC 601-1 1995
* Fullscale input
:
Ranging: Auto
Range (Mains Voltage): 90 to 264 V True RMS
Range (Accessible Voltage): 0 to 264 V True RMS
Display Units: V
Display Resolution: 1 V
Accuracy: ± 2 % of reading, ± 2 V
Crest Factor: <3
Frequency Response: DC to 1000Hz (-3dB point)
DC Input Impedance: 1 MΩ
:
Ranging: Auto
Display Units: Ω
Range: 0 to 1.999 Ω
Display Resolution: 0.001 Ω
Accuracy: ± 2 % of reading, ± 5 mΩ
Current Source Amplitude: 1 A DC (± 10 %)
:
Ranging: Auto
Display Units: MΩ
Range: 0.5 MΩ to 100 MΩ
Display Resolution: 0.1 MΩ
Accuracy: 0.5 MΩ to 20 MΩ, ± 2 % of reading, ± 200 kΩ;
above 20 MΩ, ± 5 % of reading,, ± 200 kΩ
Voltage Source Amplitude: 500 V DC (± 10 %)
:
Ranging: Auto
Display Units: A
Range: 10 A to 10000 A
Display Resolution: 10 A
Accuracy: ± 2 % of full scale
:
Voltage: >= 110 % of input Mains voltage (at no load)
Phase: In-phase or 180 out-of-phase with Mains voltage