Error Codes; Power-Up Error Codes; Start Up Diagnostic Error Codes; Sart Up Diagnostice Errors - Fluke ESA601 Operator's Manual

Electrical safety analyzer
Table of Contents

Advertisement

Error Codes

The ESA601 performs self-diagnostic tests during power-up and startup. If a problem is
detected during these times, an error code will be displayed.

Power-up Error Codes

A power-up error is indicated if the ESA601 has the power LED on, fault LED on, blank
display and continuously outputs one of the following codes to the Serial Transmit line. If
any of the power-up error codes occur, contact your local Fluke service center to arrange
repair of the ESA601.
0x01
0x05
0x09

Start Up Diagnostic Error Codes

If a fault is found at startup, a red light may appear in the
ESA601 displays one of the following error codes:
Code
Err0
Err1
Err2
Err3
Err4
Err5
Table 7-1. Power Up Error Codes
Error Code
Table 7-2. Sart Up Diagnostice Errors
Description
EEPROM checksum error.
- The unit probably needs calibration
Input AC is out of range.
(Turns ON the POWER FAULT LED).
-The meter may be out of calibration.
Physical Earth to Neutral voltage is incorrect.
(Turns ON the POWER FAULT LED.)
This may mean that:
- AC Hot and Neutral are reversed.
- There is a bad connection to Earth Ground.
- The meter may need calibration.
Error during SPI Bus Initialization.
-This is a fatal error.
Error in reading back SPI Bus data from the
power control board.
-This is a fatal error.
Continuous input overload.
Operator Messages and Service
Meaning
INSTRUCTION-TEST ERROR
RAM-TEST ERROR
FLASH-CHECKSUM TEST ERROR
POWER FAULT LED
User Action
To continue, hold down the
OHMS-FUNCTION SWITCH in
the OFFSET/ZERO 0 position
To continue, hold down the
OHMS-FUNCTION SWITCH in
the OFFSET/ZERO 0 position.
To continue, hold down the
OHMS-FUNCTION SWITCH in
the OFFSET/ZERO 0 position.
Contact Fluke Service
Department.
Contact Fluke Service
Department.
To reset this error, remove
inputs to the ESA601. Then,
power OFF and back ON.