Dual-Lead-Leakage-Current Test

The Dual-Lead-Leakage-Current test measures the current flowing from the RED signal
connection to the BLACK signal connection. Figure 5-12 shows the electrical
connections between the ESA601 and the DUT. DUT power is on. The table below the
diagram indicates the position of switches in the diagram during the test.
The test can be performed with
POLARITY SWITCH
S5
L1
L2
MAINS
OPEN
L2
S1
EARTH
OPEN
S3
TEST LEAD
PE
Figure 5-12. Dual-Lead-Leakage-Current Test Configuration
Action
1. Connect the RED and BLACK test leads to the RED and BLACK test jacks on the
ESA601.
2. Position the
3. Connect the RED test lead to the DUT enclosure; connect the DUT power cord to the
test receptacle on the ESA601.
4. Connect the BLACK test lead to the protective earth or any other test point.
5. After the meter settles, read the display (current flowing between the RED and
BLACK test leads), and/or press the
6. Repeat steps 4 and 5 for any other exposed conductive part of the DUT.
NORMAL
, single-fault condition
DUT_L1
DUT_L2
S2
DUT_PE
BLACK
RED
TERMINAL
TERMINAL
MD
A
SWITCH
DIAGRAM
ESA601
REFERENCE
NAME
S1
NEUTRAL
S2
POLARITY
S3
EARTH
S5
(N/A)
-
FUNCTION
SELECTION KNOB TO DUAL LEAD LEAKAGE
Performing Electrical-Safety Tests
and
Mains supply polarity using the
REVERSE
2 using the
OPEN L
NEUTRAL SWITCH
DEVICE UNDER TEST
CONDUCTIVE PART
ACTION
VARIABLE
VARIABLE
OPEN
to print the reading.
PRINT SWITCH
Local Control
.
APPLIED
PART
FE
TEST LEAD
ayr16f.eps
.
5
5-29