External Triggering - Agilent Technologies L34410A User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

Temperature is measured using the 2–wire or 4–wire resistance function. The above tables
N O TE
for resistance apply to temperature, but all temperature measurements use autoranging.
The range in use cannot be predicted; thus, nor can the delay. Use the TRIGger:DELay?
command to query the actual trigger delay for a temperature measurement.
For capacitance measurements the automatic trigger delay is zero (capacitor discharge is
part of the measurement). The continuity and diode test functions ignore the trigger delay.

External Triggering

External triggering takes one reading (or the specified number of readings)
each time the multimeter receives a pulse on the rear–panel Ext Trig connector.
You may select whether the multimeter uses the rising edge (POS) or falling
edge (NEG) of the external trigger signal to trigger a reading (see
Slope"
a negative pulse).
• The multimeter buffers one external trigger. If the multimeter is currently
• The Trig annunciator turns on when the multimeter is waiting for an
• The rear panel VM Comp (voltmeter complete) connector provides a pulse
34410A/11A/L4411A User's Guide
on page 77). The diagram below shows the Ext Trig connector timing (for
taking a reading while another trigger occurs, that second trigger is
accepted. After the reading in progress is completed, the stored trigger is
issued.
external trigger.
after completion of each measurement. Voltmeter complete and external
trigger implement a standard hardware handshake sequence between
measurement and switching devices. The diagram below shows the VM Comp
connector timing (for a negative pulse).
Features and Functions
"Trigger
2
75

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

L34411aL4411a 1u dmm34410a34411a

Table of Contents