Configuring Ac Voltage And Current Measurements; Configuring Resistance Measurements - Agilent Technologies L34410A User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

Configuring AC Voltage and Current Measurements

For ac voltage and ac current measurements, your menu selections are:
AC FILTER, RANGE, NULL and NULL VALUE.
• AC FILTER: Allows you to select one of three choices (3 HZ : SLOW, 20 HZ :
• RANGE: Allows you to let the multimeter automatically select the range using
• NULL: Allows you to enable (ON) or disable (OFF) the null measurement
• NULL VALUE: Allows you to view and edit the null value (if enabled).

Configuring Resistance Measurements

For 2–wire and 4–wire resistance measurements, your menu selections are:
INTEGRATION, RANGE, OFFSET COMP, AUTO ZERO, NULL and NULL VALUE.
• INTEGRATION: Allows you to set the integration time for the measurement in
• RANGE: Allows you to let the multimeter automatically select the range using
• OFFSET COMP: Allows you to enable (ON) or disable (OFF) the offset
• AUTO ZERO: Allows you to enable for all readings (ON), enable for one reading
• NULL: Allows you to enable (ON) or disable (OFF) the null measurement
• NULL VALUE: Allows you to view and edit the null value (if enabled).
34410A/11A/L4411A User's Guide
MEDIUM, 200 HZ : FAST). The ac filter allows you to trade off low–frequency
bandwidth versus ac settling time.
the autoranging feature (select AUTO), or you may select a fixed range
(select MANUAL).
feature, which measures the difference between a stored null value and the
input signal.
two ways; in power–line cycles (NPLC) or in seconds (APERTURE).
the autoranging feature (select AUTO), or you may select a fixed range
(select MANUAL).
compensation feature. With offset compensation enabled, the multimeter
makes a normal resistance measurement first, followed by a second
measurement to determine any offset voltage in the input circuitry. The
resultant displayed measurement corrects for this offset. Enabling offset
compensation increases measurement time.
(ONCE), or disable (OFF) the auto zero feature, which subtracts a subsequent
zero reading from each measurement.
feature, which measures the difference between a stored null value and the
input signal.
Features and Functions
2
45

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

L34411aL4411a 1u dmm34410a34411a

Table of Contents