Resolution
Overall Bandwidth, DCV and DCI
Triggering
Timebase Resolution
Trigger Jitter
External Trigger Latency
Internal Trigger Level Accuracy
Spurious-Free Dynamic Range and SNDR
Function
DCV
DCI
[1] 10 V range specifications are valid for signals 2 V(p-p) < x(t) < 16 V(p-p)
[2] 100 µA range specifications are valid for signals 28.8 µA(p-p) < x(t) < 200 µA(p-p)
34410A/11A/L4411A User's Guide
See table on
15 kHz typical @ 20 µs aperture (–3 dB)
Pre or Post, Internal or External, Positive or Negative
19.9524 µs, 0.01% accuracy
2 µs(p-p), 20 µs(p-p) when pre-triggered
< 3 µs
1% of range
Range
Spurious-Free
Dynamic Range
100.0000 mV
1.000000 V
[1]
10.00000 V
100.0000 V
1000.000 V
[2]
100.0000 µA
1.000000 mA
10.00000 mA
100.0000 mA
1.000000 A
3.00000 A
page 126
Signal to Noise
Distortion Ratio (SNDR)
–55 dB
–75 dB
–70 dB
–75 dB
–60 dB
–50 dB
–65 dB
–45 dB
–65 dB
–65 dB
–70 dB
Specifications
40 dB
60 dB
65 dB
60 dB
55 dB
38 dB
50 dB
38 dB
50 dB
55 dB
55 dB