Configuration Settings - Agilent Technologies 1100 Series User Manual

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Using the Detector

Configuration Settings

Figure 43
If the flow cell temperature is critical for your chromatography, you may set the
N O T E
Temperature Control to off. This will lower the optical unit and flow cell temperature by
some degree C. For more details see
Manual.
74
Detector configuration settings
• Temperature Control: the optical unit
is kept on constant temperature and
improves the baseline stability in
unstable environments. See also
note below.
• UV lamp tag: for Agilent lamps with
I.D. tags. If no I.D. tag lamp is used,
detector icon will become grey (lamp
tag not ready) and analysis is
disabled.
• Cell tag: for Agilent flow cells with
I.D. tags. If no I.D. tag cell is used,
detector icon will become grey (cell
tag not ready) and analysis is
disabled.
• Help: online help.
The detector status shows "Cell-tag" in
yellow in case the flow cell with the I.D.
tag is not inserted. The detector icon is
gray and the system is not ready.
"Principle of Temperature Control"
1100 Series DAD and MWD User Manual
in the Service