6.2.5

Power supply

Specification
Device requirements:
• voltage range: 19 to 30 VDC, typ. 24 VDC
• power consumption is typically: 15W, startup: 45W
For proper operation of the device, secure adequate power supply.
Connection
Connect the power supply to HOST board terminals 25 and 26. Use the customer cable provided
with the device or use a standard cable (shielded cable is preferred) with a maximum wire
diameter of 2.5 mm
negative pole to the pin 26.
SITRANS TDL
Operating Instructions, 05/2023, A5E51429088_AA
Pressure sensor
Electrical terminals
USB port
Ethernet port for web server connection
Add-on module interface (optional)
Front panel display connection
(14AWG). Connect the positive pole of the power supply to the pin 25 and