Limit 2 (2.4.2) To Limit 10 (2.4.10); Noise Reduction (2.5); Damping Value T90 (2.5.1); Inputs And Outputs (2.6) - Siemens SITRANS TDL Operating Instructions Manual

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

9.2.4.2

Limit 2 (2.4.2) to Limit 10 (2.4.10)

Note
Parameters for "Limit 2" to "Limit 10"
Parameters for "Limit 2" to "Limit 10" are the same as those listed previously (for "Limit 1": 2.4.1.).
9.2.5

Noise reduction (2.5)

9.2.5.1

Damping value T90 (2.5.1)

Used in the damping (filtering) of process values to smooth out response to sudden changes in
measurement. Sets time after which output signal reaches 90% of final value.
Setting
Note
Minimum damping value
When damping value T90 is set to 0, a minor delay (typically one second) is expected while the
analyzer takes the measurement.
9.2.6

Inputs and outputs (2.6)

9.2.6.1

Current input (2.6.1)

Displays current input levels.
Note
Current input linked to process parameter
If a current input is linked to a process parameter, the corresponding unit is shown.
Current input 1 (2.6.1.1)
Lower range value (2.6.1.1.1)
Sets process value that corresponds to 4 mA loop current.
SITRANS TDL
Operating Instructions, 05/2023, A5E51429088_AA
0.0 ... 100.0
Parameter assignment
9.2 Setup (2)