Commissioning
7.4 System settings
Optimize scanning frequency
The correct signal frequency must now be set ("Horizontal lines. (total)").
1. In order to set "Horizontal lines (total)" correctly, it is best to use an SMPTE and a pixel
on/pixel off test pattern. If the test patterns are not available, you can use an image with
clearly defined frame and a line written "lllllll" for the setting. Optimize the signal
frequency such that no interferences are present in the picture. If the interval between the
interferences becomes larger, and these therefore also become fewer, you are moving
the slider in the correct direction.
74
NOTICE
If the "Reset" setting is selected, all learned values are deactivated in the current
Force Mode window.
SCD 19102 / SMD 19102
Instruction Manual, 10/2011, A5E01588521A-004