Error Check

Should error be found in this instrument, try corrective actions shown in the following table. If this does
not help, this instrument has possibly been broken. Please contact the nearest Konica Minolta-authorized
service facility with indicated error number.
Error
Symptom
No.
1
Power does not turn
on.
Does not normally
start up.
2
Time for machine
initialization after
switched on varies.
3
PC cannot
communicate
with CA-2500.
Item to Check
Has AC adapter been plugged in
outlet?
Has AC adapter been connected
with body?
Have you connected correct AC
adapter?
Is the AC power within rating?
It could vary depending on the
machine position when switched
on. It should not be regarded as
failure if it takes only a few
seconds.
Check whether PC correctly
recognizes CA-2500.
Check whether device driver normally
operates with device
driver manager.
Has USB cable been connected
securely?
Check the standard for USB.
29
Corrective Action
Connect AC adapter.
Connect AC adapter.
Use our specified standard accessory AC
adapter and power cord or optionally
offered AC adapter AC-A312.
* When you disconnect a wrong adapter
and reconnect an AC-A312 adapter, wait
at least 5 minutes before turning the
power switch on after turning it off.
Be sure the AC power is within ±10% of the
nominal voltage.
If long time is needed, switch power off and
contact the nearest Konica Minolta-
authorized service facility.
PC could not properly recognize this instrument
depending on PC type or how
CA-2500 was started up. Restart PC,
software, and CA-2500 according to each
instruction.
If not, uninstall software once for reinstallation.
See the instruction manual for CA-S25w.
Connect this instrument and PC securely.
Cannot use with those supporting except for
USB2.0. Use PC with which USB2.0 is usable.