Starting Up and Ending System

Starting up

Switch power on as instructed. If not, erroneous operation could occur.
1.
Switch power on for body. (Press [ | ] side.)
l Motor sounds indicating controller inside machine starts initialization.
l Fan inside body starts to run.
l 1 5-minute warm-up at least (to leave in this state after switching on.) is recommended for accurate
measurement.
2.
Start up Windows by turning on controlling PC.
3.
Connect PC and CA-2500 with USB cable.
l P C recognizes that CA-2500 has been connected.
4.
Select "CA-S25w" from Start → [Program].
l C A-S25w starts to run.

Connection recognition error

Sometimes, PC fails to recognize its connection with CA-2500 if CA-2500 was switched on after connected with
USB cable to PC which had already started up. Such condition could not allow control of CA-2500 from software.
Start up this instrument according to the procedure as above again.
l In case that PC has already started, unplug USB cable once and then connect after switching CA-2500
on.
l Switch CA-2500 on before PC if USB cable cannot be unplugged for some reason concerning setting.
Should connection error occur, unplug USB cable once and connect again. Then, both can start up
normally.

Ending

1.
Select "End CA-S25w" from "File" of menu bar.
l C A-S25w ends.
l Y ou can finish through standard application operation of OS.
2.
Finish Windows and switch off PC.
3.
Switch off body. (Press [ ] side.)
22