Swann AMU-II ph/Redox Operator's Manual page 74

Table of Contents

Advertisement

AMU-II pH/Redox
Program List and Explanations
5.3.4.4
5.3.4.5
5.4 Miscellaneous
5.4.1
5.4.2
5.4.3
72
Fault:
No message is issued in pending error list and the
No:
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Message E024 is issued and stored in the message
Yes:
list. The Alarm relay closes when input is active.
Delay: Time which the instrument waits after the input is deactivated
before returning to normal operation.
Range: 0–6'000 sec
Language: Set the desired language.
Language
German
English
French
Spanish
Set defaults: Reset the instrument to factory default values in three
different ways:
Set defaults
no
Calibration
In parts
Completely
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All other
values are set back to default values.
 Completely: Sets back all values including communication
parameters.
Load Firmware: Firmware updates should be done by instructed ser-
vice personnel only.
Load Firmware
no
yes
A-96.150.441 / 230523