Swann AMU-II ph/Redox Operator's Manual page 58

Table of Contents

Advertisement

AMU-II pH/Redox
Program List and Explanations
2.2.2
2.2.2.1
2.2.3
2.3 Sample
2.3.301
2.3.311
2.3.310
2.4 I/O State
2.4.1
56
Miscellaneous:
Case Temp: Shows the actual temperature in °C inside the transmit-
ter.
QA History: Shows the QA values of the last quality assurance pro-
cedures:
Number: Each QA measurement is assigned to a number.
Date, Time: Date and time of the assigned number.
Deviation pH: The pH deviation of the on-line instrument.
Deviation Temperature: The temperature deviation of the on-line
instrument.
QA Check successful: <Yes> or <No>, information whether the
quality assurance was successful or not.
If <Flow measurement> = None and <Temp. Sensor> = no
Sample ID: Shows the assigned sample identification. This
identification is defined by the user to identify the location of the
sample.
Temperature: Shows the predefined comp. temperature in °C.
If <Flow measurement> = Q-Flow and <Temp. Sensor> = no
Sample ID: Same as 2.3.301
Temperature: Same as 2.3.301
Sample flow: Shows the actual sample flow in l/h
(Raw value) in Hz
If <Flow measurement> = Q-Flow and <Temp. Sensor> = yes
Sample ID: Same as 2.3.301
Temperature: Shows the actual temperature in °C
(Pt1000) raw value in Ohm
Sample flow: Shows the actual sample flow in l/h
(Raw value) in Hz
Shows the actual status of all in- and outputs.
Alarm Relay:
Relay 1 and 2:
Input:
Signal Output 1 and 2:
Active or inactive
Active or inactive
Open or closed
Actual current in mA
A-96.150.441 / 230523