AMU-II pH/Redox
Instrument Setup
Set buffer
values
A-96.150.441 / 230523
Please note that this list is only valid for Swan buffers. If you use dif-
ferent buffers please ask the manufacturer.
The temperature curves for the buffer solutions
standard 1 (pH7) and
standard 2 (pH9)
are already implemented in the transmitter firmware. To program the
temperature curve for the buffer solution pH4 overwrite standard 2.
Temperature
Buffer value at 0 °C
Buffer value at 5 °C
Buffer value at 10 °C
Buffer value at 15 °C
Buffer value at 20 °C
Buffer value at 25 °C
Buffer value at 30 °C
Buffer value at 35 °C
Buffer value at 40 °C
Buffer value at 50 °C
Buffer value at 60 °C
Value pH7
Value pH9
7.13
9.24
7.07
9.16
7.05
9.11
7.02
9.05
7.00
9.00
6.98
8.95
6.97
8.91
6.96
8.88
6.95
8.85
6.95
8.79
Value pH4
3.99
3.99
3.99
3.99
4.01
4.01
4.03
4.05
4.09
31