Swann AMU-II ph/Redox Operator's Manual page 11

Table of Contents

Advertisement

AMU-II pH/Redox
Product Description
ORP Electrode
Reference
electrode
Consumables
A-96.150.441 / 230523
For the AMU-II pH/Redox four types of redox (ORP) electrodes are
available.
 The Swansensor redox (ORP) Standard is a combined gel
electrode for application in drinking water and swimming pools.
Gel electrodes can not be filled again and have a limited life time.
 The Swansensor redox (ORP) SI is a combined electrode with
liquid electrolyte (KCl) for the measurement of redox (ORP) value
in power plants.
 The Swansensor redox (ORP) AY is a combined gel electrode for
application in waste water due to additional salt supplies.
 The Swansensor ORP FL for the measurement of the redox
potential in high purity water. This sensor can only be used in
combination with Swansensor Reference FL, A-87.860.100.
Swansensor Reference FL, reference electrode for Swansensor pH
FL or Swansensor Redox FL
One 200 ml bottle of 3.5 M KCl lasts for 1 month.
9