Sensor installation
Fig. 52: Avoid: Slope and mismatch
Fig. 53: Installation to be avoided: Fixing the sensor
Fig. 54: Recommended installation: use the piping to support the sensor
1
2

12.4.1.1 Avoiding creation of noise

Zero point stability is a prerequisite for exact mass flow measurement. Insufficient
installation may lead to mechanical tensions or flow noise which impact zero point
stability.
Countermeasures to help avoid noise creation:
Avoid fixing anything directly to the sensor. Doing so may result in increased
deviations.
1
1
2
1
1
2
Pipe
Sensor
Secure pipes before installing the flow meter.
Avoid damaging the process connections.
Flush new pipes before installing the flow meter to remove foreign matter, such as
shavings or other residues.
Support sensor weight by using soft coupling (silicone or other types of cushioning
materials).
Avoid bending or tensioning the sensor while aligning the pipe.
Avoid reductions or expansions in pipe directly up- or downstream of flow meter.
Avoid placing control valves, apertures or other devices generating noise near the
sensor.
A5E52748387-AA, 2023/09
FC7x0
Installation
91 / 180