System configuration and operation

15.4.3.3 Setting the bus address

This section describes the procedure to set bus address in the transmitter. Every device
in PROFIBUS PA must be assigned a unique address in the range of 0(0x00) to
126(0x7e). If it is not specified at the time of order, 126(0x7e) is the factory default. Do
not change to 0, 1 or 2 as these are used by master devices. There are two ways to set
bus address: by using communication tools or by hardware switch in device.
Hardware Address
Device address Mark
Switch
Hardware
(ADDRESS: 7)
Software
Numbers of these address switches show the square of each number.
Hardware Address
Switch
Example: If only the address SW-6 is 1, address shows 64.
(ADDRESS: 0 – 6)
1 * 2
Address has to be set always between 0 and 126. If address switch is set to 127, the
address is automatically converted to 126. Hardware Address Switch change has to be
performed to reflect hardware switch status after powering off a device.
Device address can be set by PROFIBUS PA software tool if hardware address setting is
Software address
not active. For further details, see applicable Function Manual (FM) A5E52748624.
setting
142 / 180
Description
"Hardware Address Switch" settings between 0 and 6 are
1
device addresses.
Parameter setting is device address.
0
Factory setting.
6
5
4
3
+ 0 * 2
+ 0 * 2
+ 0 * 2
A5E52748387-AA, 2023/09
2
1
0
+ 0 * 2
+ 0 * 2
+ 0 * 2
Advanced settings