Disconnect Compact Type Sensor - Siemens SITRANS FC720 Operating Instructions Manual

Coriolis c
Table of Contents

Advertisement

FC7x0
Sensor replacement
1. Open the terminal cover.
2. Disconnect all cables with the operating tool.
3. Remove the cable gland and cable.
4. Disconnect the sensor potential equalization.

18.1.3 Disconnect compact type sensor

Life-threatening injuries from electric shock
WARNING
▶ Switch off power supply and communication.
▶ Secure against inadvertent switch-on.
After power down and breaking all power cable connections from L/+ N/- and grounding
Disconnecting
for the power circuit (except potential equalization connection of external grounding
power,
terminal) the sensor has to be disconnected by the following procedure:
communication
and grounding
1. Remove locking screw of the "Terminal box power and I/O connection" and remove
2. Disconnect the communication and I/O cables (avoid short circuit!).
3. Remove the cable gland and cable.
4. Disconnect the power cables from L/+ and N/- "Power supply connection terminals".
5. Remove "Grounding screw for connecting grounding conductor".
6. Remove potential equalization from "Grounding terminal for potential equalization" (if
152 / 180
the back cover.
used).
A5E52748387-AA, 2023/09
Operating tool
Grounding screw
Dismantling of the defective sensor

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fc740

Table of Contents