5560A/5550A/5540A
Service Manual
To calibrate the Capacitance function for values <1 mF:
1. Remove the Guard Ground strap on the Product.
2. Set the LCR meter for Cp Mode, Level 1V, Freq 1 kHz, Speed Slow.
3. Complete the OPEN/SHORT compensation of the LCR meter and lead set for two-wire
measurement that includes the Y adapter and single banana plugs. Do not connect a strap
between the Guard and Ground terminals of the DUT for this function.
4. Connect the LCR meter to the Product for two wire capacitance measurements.
5. From step 1, Table 18, set the LCR meter for the recommended frequency (the Product can be
verified at frequencies other than specified, but the measured output varies within the specifications
as a function of the frequency).
6. Select appropriate mode (Cp or Cs).
7. Output from the Product. Measure with the LCR meter and record in Table 18.
Table 18. Calibration Steps for Capacitance with LCR Meter
Step Product Range Product Output Test Frequency Calculated Product Output
1
1.2 nF
2
1.2 nF
3
12 nF
4
12 nF
5
120 nF
6
120 nF
7
120 nF
8
120 nF
9
120 nF
10
120 nF
11
1.2 µF
12
1.2 µF
13
12 µF
14
12 µF
15
120 µF
16
120 µF
17
1.2 mF
54
0.2 nF
1000 Hz
1 nF
1000 Hz
1.21 nF
1000 Hz
10 nF
1000 Hz
12.1 nF
610 Hz
20 nF
610 Hz
30 nF
610 Hz
50 nF
610 Hz
100 nF
610 Hz
120 nF
610 Hz
0.121 µF
100 Hz
1 µF
100 Hz
1.21 µF
80 Hz
10 µF
80 Hz
12.1 µF
20 Hz
100 µF
20 Hz
0.121 mF
5 Hz
Limits
±2.2 pF
±3.1 pF
±6.3 pF
±16 pF
±45 pF
±54 pF
±66 pF
±90 pF
±150 pF
±174 pF
±445 pF
±1.50 nF
±4.45 nF
±15.0 nF
±41 nF
±155 nF
±516 nF