Inductance (5550A and 5560A)

Table
20
lists the required equipment.
Table 20. Required Equipment for Inductance Calibration
Equipment Description
LCR Meter
BNC (M) to BNC (M) 4W lead set for LCR
BNC (F) To Single Banana Plug
To calibrate the Inductance function:
1.
Remove the Guard Ground strap on the Product.
2. Set the LCR meter for Ls Mode, Level 1V, Freq 1 kHz, Speed Slow.
3. Complete OPEN/SHORT compensation of the LCR meter and lead set including the banana to bnc
plugs. Do not connect a strap between the Guard and Ground terminals of the DUT for this
function.
4. Connect the LCR meter to the product for four-wire inductance measurements.
5. Starting at step 1, Table 21, set the LCR meter for the recommended frequency (the Product can
be verified at frequencies other than the specified, but the measured output varies within the
specifications as a function of the frequency).
6. Select the appropriate mode (Ls or Lp).
7. Output each value from the Product. Measure with the LCR meter and record in Table 21.
Table 21. Calibration Steps for Inductance with LCR Meter
Step Product Range Product Output Test Frequency Calculated Product Output
1
120 µH
2
120 µH
3
1.2 mH
4
1.2 mH
5
12 mH
6
12 mH
7
12 mH
8
12 mH
9
12 mH
10
12 mH
15 µH
1000 Hz
100 µH
1000 Hz
121 µH
1000 Hz
1 mH
1000 Hz
1.21 mH
110 Hz
2 mH
110 Hz
3 mH
110 Hz
5 mH
110 Hz
10 mH
110 Hz
12 mH
110 Hz
Calibration Verification of the Product
Manufacture/Model/PN
Hioki IM3533
Various
Pomona Model 1894
Calibrator
Quantity
1
1
4
Limits
±227 nH
±380 nH
±1.13 µH
±2.1 µH
±11.3 µH
±12.2 µH
±13.3 µH
±15.5 µH
±21.0 µH
±23.2 µH
57