12. Starting with step 40, make a measurement with the 5790B. Record the measurement in the
Product Output column of Table 12.
13. Continue with the rest of the steps for this connection (steps 40 to 121).
14. With the entire table populated, evaluate each step for compliance to limits. If Guardbanding is to
be applied create a table from the example in this manual, where the acceptance limits are properly
modified by your Guardbanding method.
Step
Product Range
1
12 mV
2
12 mV
3
12 mV
40
12 mV
41
12 mV
42
12 mV
43
12 mV
44
12 mV
45
12 mV
4
12 mV
5
12 mV
Figure 18. Volts AC Calibration
Table 12. Calibration Steps for Volts AC
Product Output Product Frequency
1 mV
3 Hz
1 mV
5 Hz
1 mV
10 Hz
1 mV
1 kHz
1 mV
20 kHz
1 mV
50 kHz
1 mV
100 kHz
1 mV
300 kHz
1 mV
500 kHz
10 mV
3 Hz
10 mV
5 Hz
Calibration Verification of the Product
5590B
Calculated
Product
Output
±9.00 µV
±7.70 µV
±6.12 µV
±6.12 µV
±6.12 µV
±6.30 µV
±16.2 µV
±36.4 µV
±36.4 µV
±27.0 µV
±14.0 µV
Calibrator
Limits
35