Expert 2" Menu - Eizo CuratOR EX190 Instructions For Use Manual

1mp 19" lcd monitor
Table of Contents

Advertisement

7.13 "Expert 2" menu

Note
Opening and closing the "Expert 2" menu
• Open the "Expert" menu by pressing keys 1 and 3 in the "Info" menu in sequence.
• Navigate with operation key 3 to the "Expert 2" menu.
• The "Expert 2" menu closes automatically upon exiting the OSD.
Function
Timing change
delay adjust-
ment
CuratOR EX190
Instructions for Use, 06/2022
Values
Description
Active
When "Active" is selected, the timing change delay can be
set.
Inactive
The following four parameters are relevant to this function:
Default: Inactive
Changing one of these parameters corresponds to a tim-
ing change, and the monitor begins to resynchronize
("Auto in progress" display). To prevent this from happen-
ing as a result of each and every signal disturbance, the
value representing the permissible number of faulty or
changed borders can be increased in the case of unstable
signal sources using the "Signal change delay" function.
This is required in the case of unstable signal sources, for
example.
Note: The disadvantage of the function is a delay of sev-
eral milliseconds even for a desired timing change. The
higher the value of the function, the greater the delay.
H frequency
V frequency
V total
Interlaced/non Interlaced
Operation
7.13 "Expert 2" menu