6.5.3 Entering timing data

If you know or have already determined the timing data, you can enter it directly in the OSD
"Force Mode" menu.
Entering known timing data
1. Open a test pattern with a clearly defined border (e.g. SMPTE image).
2. Open the "Force Mode" menu is the OSD.
3. Use the keys to enter the timing data in the input fields.
4. Set the "Mode" to "Active".
5. Apply the entered data by selecting "Execute function..."
Note
Always save the data with "Execute function...".
The timing data are only entered into memory with the "Execute function..." command, and
the image is then displayed with the new parameters.
CuratOR EX190
Instructions for Use, 06/2022
6.5 Adapting the monitor using Force Mode
Commissioning