Eclipse 1000 User Manual page 46

Hide thumbs Also See for 1000:
Table of Contents

Advertisement

If tubing becomes disconnected during operation, ECLIPSE does
WARNING
not alarm. Check flow at the cannula to verify oxygen delivery.
DO NOT tamper with, disassemble, crush or heat the Power Cartridge
WARNING
above 140° F (60° C). The Power Cartridge may present a risk of fire
or explosion and will void the warranty.
DO NOT expose the AC and DC Power Supplies to water. Electrical
WARNING
shock or serious injury may result.
DO NOT disassemble the ECLIPSE or attempt to perform any routine
WARNING
maintenance other than what is described in the Patient Cleaning and
Care section below. Disassembling the oxygen system may create a
hazard of electrical shock and void your warranty. Contact your
dealer/provider for any maintenance required.
Unplug Power Cords, AC or DC Power Supplies before cleaning
WARNING
the exterior cabinet. DO NOT use denatured alcohol or apply liquid
spray or aerosol cleaners.
DO NOT tamper with or try to repair the Power Cartridge. No
WARNING
serviceable parts inside.
Exposing the Power Cartridge to water or other liquids may cause
WARNING
personal injury.
DO NOT drip, spill or expose the ECLIPSE to water during portable
WARNING
use. The ECLIPSE enclosure does not protect against the harmful
effects of liquid ingress. Electrical shock or damage to the unit
may result.
44 | ECLIPSE
TM
Model 1000
User Manual