The filter should be free of liquid water and dry to the touch before
CAUTION
reinstallation, as excess moisture may impair the proper operation of
the oxygen system.
When the filter is removed, the Service Port is exposed. DO NOT
CAUTION
attach any cables to this Service Port. It is for Service ONLY .
DO NOT poke objects through the vent holes.
CAUTION
The air inlet filter should be replaced as needed. If the oxygen system is used in a
dusty environment, the filter may need to be replaced more often. You should not
operate the ECLIPSE without a filter installed for longer than 30 minutes.
Cabinet, AC Power Supply, DC Power Supply and Power Cord:
ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection
activity. Use a damp (not soaking wet) cloth or sponge and a mild detergent solution
to clean the cabinet and power supplies. To disinfect the ECLIPSE use Lysol
disinfectant. Proceed as directed by the manufacturer.
Unplug Power Cords, AC or DC Power Supplies before cleaning the
WARNING
exterior cabinet. DO NOT use denatured alcohol or apply liquid spray
or aerosol cleaners.
If used, clean your humidifier daily to reduce possibility of contamination.
Humidifier:
Replace humidifier monthly or as recommended by the manufacturer or your
dealer/provider.
Before cleaning the cart, remove the ECLIPSE. Use a damp (not soaking wet)
Cart:
cloth or sponge and mild detergent solution to clean the Cart.
36 | ECLIPSE
TM
Model 1000