Pulse Flow Mode Operation
I
• Pulse Flow Mode should only be used under the direction of a qualified
clinician or physician.
• When operating in Pulse Flow Mode, an alarm will beep after 15 seconds
if inspiration is not detected. If another 15 seconds elapses and no
inspiration is detected, the unit will switch to Continuous Flow Mode
at the last Continuous Flow Mode rate setting used.
Pulse Flow settings should be determined for each patient individually.
WARNING
Settings from Continuous Flow applications may not be applicable to
Pulse Flow Mode.
When operating in Pulse Flow Mode, to ensure adequate breath
WARNING
detection, the length of the tubing/cannula must not exceed
7 feet (2.1m).
As with all conserving devices, the ECLIPSE may not be able to detect
WARNING
some respiratory efforts.
STEP 5: ADJUST FLOW RATE TO PRESCRIBED LEVEL
Using the
Flow Rate Setting buttons
I
the prescribed setting supplied by your physician.
DO NOT change from prescribed flow rate setting without consulting
WARNING
with a qualified clinician or physician.
28 | ECLIPSE
TM
Model 1000
User Manual
, adjust the flow rate to